To produce x-rays, a cloud of electrons are necessary and the cathode must have a strong positive charge.
False
True
2
Reducing the amount of scatter radiation reaching the IR results in:
Increased exposure to the IR
Lower radiographic contrast
Higher radiographic contrast
No change in radiographic contrast
3
Increasing collimation (decreasing the field size) results in:
Decreased scatter production and lower contrast.
Decreased scatter production and higher contrast.
Increased scatter production and lower contrast.
Increased scatter production and higher contrast.
4
A flat piece of lead with a hole in it that has a flange directed toward the patient is a(n):
Grid
Collimator
Cone or Cylinder
Filter
5
The motor that turns the anode operates through what?
Alternating current
Static electricity
Direct current
Mutual induction
6
Adding a grid or increasing grid ratio results in an increase in patient dose.
False
True
7
Absorption of transmitted photons by a grid caused by misalignment is:
Grid cutoff
Scatter radiation
Quantum noise
Focal spot blur
8
The step-down transformer is found on the:
filament circuit
none of these
primary circuit
secondary circuit
9
The formula for the grid conversion factor is:
GCF = kVp with grid/kVp without grid.
GCF = mAs with grid/mAs without grid.
GCF = mAs without grid/mAs with grid.
GCF = SID with grid/SID without grid.
10
Grids absorb scatter radiation because the scattered photons:
Travel in the same direction as the transmitted photons
Travel at an angle to the transmitted photons
Travel parallel to the transmitted photons
Are absorbed by the patient
11
Grids are typically used when:
Part measures more than 10 cm and at least 60 kVp needed for the exam only.
Low contrast images are desired
Less than 60 kVp is needed for the exam
The part measures less than 10 cm
12
When imaging the pelvis, changing from 14 × 17 collimation to 8 × 10 collimation requires:
An increase in kVp.
No change in mAs.
A decrease in mAs.
An increase in mAs.
13
When using automatic collimation, collimating to the anatomy of interest is an unnecessary waste of time.
False
True
14
What does the cathode include?
Filament only
Induction Motor only
Focusing Cup and Filaments
Target and Induction Motor
15
The focusing cup is able to keep the electrons together because:
Opposite charges attract
There's a vacuum inside the tube
The anode generates a magnetic field
Like charges repel
16
When the x-ray beam is angled across the lead strips you have _________ cutoff.
Off-axis
Off-center
Off-level
Off-focus
17
The rate of flow of electrons passing through the x-ray tube during an exposure is:
tube current
filament current
Hz
kVp
18
With a rotating anode, the focal spot becomes a focal:
Point
Circle
Track
Line
19
What component of the X-ray tube is responsible for producing electrons?
Filament
Focusing cup
Anode
Cathode
20
For radiographic procedures, scatter radiation is primarily the result of:
Photoelectric effect
Compton interaction
Pair production
Coherent scattering
21
The timer circuit is found on the:
primary circuit
filament circuit
secondary circuit
none of these
22
Which of the following is true regarding the effect of mAs on the brightness of a digital image?
mAs decreases brightness
mAs does not affect brightness in a digital image
Decreasing mAs increases brightness
Increasing mAs increases brightness
23
The unrestricted primary x-ray beam shape is:
Rectangular
Round
Square
Oval
24
What is the relationship between kVp and subject contrast?
High kVp creates high subject contrast.
High kVp creates low subject contrast.
Low kVp creates high subject contrast.
Low kVp creates low subject contrast.
25
To compensate for the reduction of radiation reaching the IR when a grid is used, ___________ is typically increased.
kVp
Exposure time
SID
mAs
26
The number of lead lines per inch or centimeter is the grid:
Grid factor
Frequency
Density
Ratio
27
In addition to the oil surrounding the tube, mechanical fans help to eliminate the heat produced during x-ray production
False
True
28
The devices that change alternating to direct current are the:
rectifiers
filaments
transformers
rheostats
29
The primary side of the high-voltage transformer is found on the:
primary circuit
secondary circuit
filament circuit
none of these
30
A grid with lead strips that run parallel to each other is a:
Parallel grid and nonfocused grid.
Angled grid
Crossed grid
Focused grid
31
The recommended SIDs that can be used with a focused grid is the:
Focal distance
Grid convergence
Grid ratio
Focal range
32
The beam-restriction device that results in images with the most unsharpness around the edge of the image is the:
Collimator
Aperture Diaphragm
Cone or Cylinder
Grid
33
With digital imaging, if the mAs is too high:
The image will be underexposed.
The image will appear blurry.
The image brightness will appear appropriate.
The image will be overexposed.
34
Anyone in the room during an exposure must wear a lead apron to absorb the scatter radiation produced in the:
Lights
Patient
Sun
Radiologist
35
Linear grids are the most commonly used because:
The tube can be angled in the direction of the lead lines.
They absorb more scatter radiation.
They are cheaper than other types of grids.
They are easier to manufacture.
36
The two major factors that affect the amount of scatter radiation produced and exiting the patient are:
Tube current and exposure time
SID and OID
Volume of tissue irradiated and kVp
Filtration and collimation
37
Grids absorb almost all of the scatter radiation and allow all of the transmitted photons to pass through to the IR.
True
False
38
The main power switch is found where?
secondary circuit
filament circuit
primary circuit
none of these
39
To overcome the decrease in sharpness caused by the air gap technique, an increase in SID is required.
False
True
40
Which of the following factors need to be considered when determining the grid to be used for a specific examination?
Level of contrast improvement needed.
Patient exposure.
Focal range.
All of the above
41
Higher kVp results in less absorption and more variation in the intensities exiting the patient.
False
True
42
An IR that has a grid permanently attached to its front surface is a:
Focused cassette
Nonfocused cassette
Grid holder
Grid Cassette
43
Which of the following statements is true regarding practices to minimize patient exposure?
Using a grid increases patient dose
All of these
Increasing kVp reduces patient dose
Using lower mAs decreases patient dose
44
The air gap technique is based on creating a gap by increasing the ______.
SID
mAs
Focal spot size
OID
45
To maintain the same exposure to the IR, if the SID is increased, the mAs must be:
Kept the same
Decreased
Multiplied by the square of the SID
Increased
46
The volume of tissue irradiated is determined by:
All of the above
Patient thickness.
Field size
The area of collimation.
47
For a lateral lumbar spine, placing a lead shield on the table behind the area of exposure:
Reduces the scatter reaching the IR.
Decreases patient dose.
Improves image contrast.
Increases the scatter reaching the IR.
48
The most effective and often used radiographic beam-restriction device is the:
Filter
Collimator
Grid
Cone or Cylinder
49
The effective focal spot is smaller than the actual focal spot because the face of the anode is angled.
True
False
50
The only grid cutoff error that can occur with a parallel grid is:
Off-focus
None of these
Off-level
Off-center
51
The grid is part of a device located just below the tabletop that also includes a tray to hold the IR This device is the
Collimator
Cone or Cylinder
Bucky
Filter
52
Prepping the rotor but not making an exposure on a regular basis is related to tube damage or failure.
True
False
53
Grid cutoff results in:
May increase patient exposure.
Possible increase in quantum noise.
All of the above
Reduced exposure to the IR.
54
The x-ray tube (not filaments) is found on the:
none of these
filament circuit
primary circuit
secondary circuit
55
Which of the following occurs first in the sequence of events leading to x-ray production?
Voltage becomes kilovoltage
AC is converted to DC
Anode and cathode become positively and negatively charged
Electrons flow from cathode to anode
56
AEC controls the ________________ of radiation reaching the image receptor.
quantity
energy level
type
all of these
57
Place the following events leading to the production of x-rays in order, from first to last:
1, 3, 4, 2
2, 4, 1, 3
4, 3, 2, 1
3, 1, 2, 4
58
The device that adjusts the incoming power supplied to the x-ray circuit to maintain 220 V is the:
autotransformer
circuit breaker
line compensator
timer circuit
59
The device used to absorb scatter radiation leaving the patient before it reaches the image receptor is the
Lead apron
Grid
Collimator
Filter
60
If all the convergent points were connected along the length of the grid they would form the:
Convergent line
Grid line
Focal line
Parallel line
61
The automatic collimator:
Is not commonly used
Requires manual adjustment
Cannot be adjusted
Automatically collimates to the IR size
62
What does the anode include?
Target and Induction Motor
Focusing Cup and Filament
Induction Motor only
Filament
63
The relationship between the height of the lead strips and the distance between them is the grid:
Frequency
Density
Ratio
Grid Factor
64
The purpose of the filament circuit is to:
provide direct current to the x-ray tube
boil electrons off the filament
produce high voltage
reduce the current reaching the filament
65
To compensate for some of the loss of SR due to the large OID used with the air gap technique, the ________
Exposure Time Increases
kVp Increases
SID Increases
mAs Increases
66
The field shape produced by a collimator is:
Round
Square
Oval
Rectangular
67
Which of the following is not a primary exposure technique factor?
kVp
SID
mA
Time
68
When the SID is above or below the focal range you have _________ cutoff.
Off-axis
Off-focus
Off-level
Off-center
69
Using a higher the grid ratio will:
Increase scatter clean up and decrease radiographic contrast only.
Decrease scatter clean up and decrease radiographic contrast only.
Increase scatter clean up and increase radiographic contrast only.
Decrease scatter clean up and increase radiographic contrast only.
70
The stator is made up of:
Permanent magnets
Oil-filled tubes
Bearings
Electromagnets
71
The grid is located:
Between the patient and the detector
Outside of the x-ray room
Between the patient and the IR
Between the x-ray tube and the patient
72
Place the following events leading to the production of x-rays in order, from first to last:
1, 2, 3, 4
2, 4, 3, 1
2, 1, 4, 3
4, 3, 2, I
73
The primary purpose of the step-down transformer is to:
increase the voltage going to the filaments
increase the amperage going to the filaments
decrease the amperage going to the filaments
decrease the voltage going to the filaments
74
Which of the following produces the greatest amount of scatter radiation exiting the patient?
Low mAs and large field size
Low kVp and small field size
High kVp and large field size
High mAs and small field size
75
Scatter radiation is of no value in radiographic imaging.
False
True
76
Which of the following is the formula used to determine grid ratio?
Grid ratio = distance between the lead strips × height of the lead strips
Grid ratio = distance between the lead strips/height of the lead strips
Grid ratio = height of the lead strips × distance between the lead strips
Grid ratio = height of the lead strips/distance between the lead strips
77
Place the following events leading to the production of x-rays in order, from first to last:
2, 4, 3, 1
1, 2, 3, 4
3, 4, 2, 1
2, 3, 4, 1
78
The probability of a Compton interaction occurring is related to:
The energy of the x-ray photon.
The wavelength of the x-ray photon
The atomic number of the material
The density of the material
79
The distance between the grid and the convergent line is the:
Grid frequency
Focal distance
Grid convergence
Grid ratio
80
The Moiré effect can be caused by:
Grid misalignment
Using a grid Cassette in the bucky
Using a grid with similar frequency to the CR laser scanner frequency
Both Red and Blue answer choices
81
A significant amount of collimation will have what effect on the amount of radiation:
Absorbed
Unchanged
Increased
Reduced
82
Using excessive kVp with a digital IR results in:
Increased IR exposure decreased scatter to the IR and increased contrast
Decreased IR exposure decreased scatter to the IR and decreased contrast
Increased IR exposure increased scatter to the IR and decreased contrast
Decreased IR exposure increased scatter to the IR and increased contrast
83
What is the grid ratio if the lead strips are 2.4 mm high, 0.02 mm wide, and 0.3 mm apart?
6:1
8:1
12:1
10:1
84
The grid limits the amount of scatter radiation produced in the patient.
False
True