Scintillation Detectors (Instrumentation Textbook) Flashcards | StudBud
Explore
/
Scintillation Detectors (Instrumentation Textbook)
Scintillation Detectors (Instrumentation Textbook)
61 terms
·
StudBud Administration
·
26 plays
Imported from external platform
Flashcards
61 cards
F
Practice quiz
P
Host live
L
Questions
61
Answers
1
____ is when the next pulse arrives before the pulse has returned to baseline, results in incorrect low energy
baseline shift
pulse pileup
amplification
noise
2
What can interact with a scintillation crystal?
beta particles
alpha particles
gamma ray
3
The ___ corresponds to gamma rays that have undergone photoelectric absorption in the scintillation crystal.
peak broaden
pulse
photopeak
pulse height
4
The PMT must be shielded against stray magnetic fields, as these will alter the trajectory of the electrons.
False
True
5
The total multiplication factor of the dynodes is about ___
one billion
one trillion
one million
one thousand
6
Most counts to the right of the photopeak have undergone a Compton interaction and are considered "scattered" photons
True
False
7
The well configuration is designed for external counting of radioactivity in organs such as the thyroid gland.
True
False
8
Scintillation detectors allow us to distinguish between scattered and unscattered gamma rays.
True
False
9
Calibration is needed to correlate _____ to the energy of the gamma rays being detected.
dynodes
pulse size
photons
electrons
10
Which of these do NOT contribute to peak broadening?
statistical variations in electron reproduction at the crystal
the gamma ray energy is not reproduced exactly in the detector electronics
the type of radioactivity
statistical variation in the # of electrons produced at the photocathode*
11
A scintillation detector has the ability to determine the energy of the interacting gamma ray.
False
True
12
The valence band is at a ____ energy than the conduction band.
lower
higher
13
In a probe configuration, you have a flat-field collimator to give meaningful measurements of individual organs.
True
False
14
An increasing energy resolution at quarterly checks could mean everything except:
a problem with the PHA
yellowing of the crystal
a problem with electronic noise
a cracked crystal
15
The y-axis of an energy graph represents:
voltage
pulse
the number of events w/ energy in that range
narrow range of pulse heights
16
the ___ collects all the electrons produced from the final dynode and emits an output voltage signal.
PMT
anode
cathode
PHA
17
If your readings are higher than the expected constancy reading it could be because of everything except:
contamination of detector or housing
crystal damage
incorrect source-detector geometry (closer than usual)
extraneous radiation source in detector's active volume
18
If your readings are lower than the expected constancy reading it could be because of everything except:
contamination of detector or housing
toggle switch pointing to other detector
incorrect threshold and/or window setting
incorrect high-voltage setting
19
_____ evaluates how well a given detector distinguishes between gamma rays of closely spaced energies.
energy resolution
pulses
spatial resolution
photocathodes
20
Another name for energy graph is:
peak broaden graph
voltage curve response
photopeak graph
pulse height spectrum
21
A detector's _____ characteristics determine how well that scale applies throughout the energy range.
amplifying
energy linearity
dynode
PMT
22
The thallium impurities create electron orbitals b/t the conduction and valence bands called ____
forbidden centers
gap centers
excited states
activation centers
23
The ____ counts accepted pulses and tracks time.
PMT
photocathode
scalar/timer
anode
24
A drawback of a single-channel analyzer is:
it also counts radiation events above & below the ULD & LLD
it rejects radiation events that don't fall between the LLD & ULD
25
A _____ converts scintillation photons to electrons.
crystal
PMT
anode
photocathode
26
Scintillation detectors with 1 PMT are made in 2 specific configurations:
crystal & probe
probe &. crystal
probe & well
crystal & pulse height analyzer
27
When an electron moves from the ___ band to the ____ & back down to the ground state, it releases a scintillation photon
conduction, valence
valence, conduction
28
Where must a pulse fall in relation to ULD and LLD in order to produce PHA output pulses?
in between the ULD & LLD
to the left of the LLD
below the ULD
above the LLD
29
Which instrument is NOT a scintillation detector?
well counter
thyroid probe
cutie pie
PET
30
The scintillators used in NM are inorganic ____ scintillators, often w/ small amounts of impurities.
metal alloy
crystalline
glass
noble gases
31
____ is the ability to convert radiation emissions to detected signals. Depends on composition, geometry, energy window
baseline shift
detector efficiency
pulse pileup
32
A series of ___ multiply the small # of electrons generated at the photocathode into an electronic signal.
cathodes
high voltage supplies
anodes
dynodes
33
the _____ determines the amplitude of the pulse, which correlates with the gamma ray's energy.
pulse height analyzer
preamplifier
PMT
amplifier
34
The ____ provides the proper electric field to direct the electrons toward the first dynode
focusing grid
photocathode
crystal
PMT
35
A well counter is not as sensitive as a GM survey meter.
False
True
36
A ____ boosts the strength of the PMT output signal to allow its passage through the remaining detector electronics.
preamplifier
amplifier
pulse height analyzer
37
The most common inorganic scintillator employed in NM is _____ activated sodium iodide.
Xenon
Thallium
Cesium
Potassium
38
Scintillation photons are detected by a _____ and are converted to an electronic signal.
photmultiplier tube
preamplifier
pulse height analyzer
amplifier
39
___ is the number of scintillation photons emitted per eV or radiation absorbed
scintillation light yield
counts
scintillation efficiency
TDE
40
An MCA doesn't show all radiation events on an energy graph.
False
True
41
Cs-137 emits ____ keV gamma rays.
530
140
662
842
42
The well configuration is designed to measure samples of radioactive materials in test tubes.
True
False
43
The output pulses from the PHA are the same size and are sent on to the scaler and rate meter.
False
True
44
scintillation detectors generally operate in ___ mode.
current
pulse
45
The energy resolution will be larger for lower-energy photo peaks than for high-energy photo peaks
False
True
46
We quantify the energy resolution by calculating the
Full Width Half Maximum
short width half maximum
Half width full maximum
total width total maximum
47
A vacuum must be maintained inside the PMT so that no electrons are lost due to interaction w/ any gas molecules.
True
False
48
when pulses pile on top of one another & recorded as a single event. Can cause scattered counts to be registered in PHA
amplification
pulse pileup
baseline shift
noise
49
Which energy is always used to compare energy resolution between detectors because of only having a 6-8% FWHM?
Tc-99m
Cs-137
50
___ is any material that can release a photon in the UV or visible-light range.
scintillator
gamma ray
photoelectron
HVL
51
Which is NOT a type of pulse height analyzer?
single-channel
multichannel
preamplifier
52
_____ allows us for better counting of photons at close energies
energy spectrum
photopeak
ULD
peak broadening
53
The ULD and LLD are set by the operator & are related to the ___ of the photons being measured.
energy
speed
mass
velocity
54
The ____ of the PMT determines the potential difference b/w the dynodes & the final size of the PMT output pulse.
high voltage supply
gasses
scintillation light yield
amount of dynodes
55
Scintillation photons are converted to an electronic signal by the ___
crystal
anode
photocathode
PMT
56
The X-axis of a energy graph represents:
voltage
pulses
a narrow range of pulse heights
counts
57
We deal with ____ by setting the detector to look at a range of energies around the photo peak.
pulse height
PMT
peak broadening
counts
58
Most ___'s work by assigning the event to one of a series of predefined bins or channels based on the signal's size
SA
anode
amplifier
MCA
59
We set up an ____ to include most of the photo peak and exclude everything else.
dynode
scintillation photon
energy window
pulse height
60
A 1% change in the high voltage results in a 10% change in the ____ size.
current
electron
pulse
energy
61
An ____ amplifies & shapes the preamp output signal by a variable amount & is controlled by course & fine grain controls
preamplifier
pulse height analyzer
PMT
amplifier