RADR 2205 PRI II - Comprehensive Review Flashcards | StudBud
Explore
/
RADR 2205 PRI II - Comprehensive Review
RADR 2205 PRI II - Comprehensive Review
99 terms
·
StudBud Administration
·
99 plays
Imported from external platform
Flashcards
99 cards
F
Practice quiz
P
Host live
L
Questions
99
Answers
1
A periodic backup of archived files
Image Recovery Plan
Disaster Recovery Plan
RIS Recovery Plan
Required Backup Plan
2
What type of artifact is this?
Dropped/dead pixels
Grid cut-off
Electronic interference
Grid lines
3
The light field/x-ray field congruence must be within ____ of the SID
5%
3%
1%
2%
4
is a preprocessing software correction that is performed to equalize the response of each pixel to a uniform x-ray beam.
Flat fielding
Interpolation
Pixel averaging
Histogram analysis
5
The ·Control panel must indicate when the tube is energized, with light and sound
False
True
6
Pulsed Fluoroscopy is intermittent radiation with high mA values and reduces exposure to patients and personnel
False
True
7
The goal of IHE is to improve the interoperability of healthcare information systems by promoting established standards
True
False
8
Level 2 QC is:
Invasive and complex
Invasive and simple
Noninvasive and simple
Noninvasive and complex
9
Pulsed Fluoroscopy is constant radiation exposure and can produce images with quantum mottle and low SNR
True
False
10
A universally accepted standard for exchanging medical imaging via display workstations, archive, & imaging modalities
PACS
DICOM
MIMPS
HL7
11
What should the exposure rate above the tower not exceed when the tower is at maximum SID with shutters fully open?
5 mR/h
2 mR/h
10 mR/h
20 mR/h
12
The input phosphor of an image intensifier tube converts a(n) _____ image into a(n) _____ image.
light; electron
x-ray; light
light; electron
x-ray; electron
13
What is Health Level 7 (HL7)'s purpose?
professional society for Imaging Informatics professionals
Set standards on how information systems share patient data
physical security measure in hospitals
encrypts patient data
14
Benefits of digital PACS images do not generally include
Convenient storage
computer aided diagnosis
improved spatial resolution
image transmission
15
When testing AEC detectors, the detectors must produce pixel brightness within ___ of each other
50%
20%
30%
40%
16
This level of quality uses statistical data to measure outcomes & considers all factors that can influence them
Actual quality
Expected quality
Perceived quality
17
Maximum variance for mAs Reciprocity is within ____
20%
10%
15%
5%
18
CQI targets the process in which employees operate, a process is defined as:
a data driven approach that helps management make decisions to improve
a systematic collection and evaluation of data
a series of steps to achieve a desired outcome
a measurement of the level of quality with no effort to change or improve
19
A display workstation is:
storage device that keeps all digital image data
any computer used to view a digital image
a system constructed to transmit; receive, and store image data
a device that is used to capture digital medical images
20
What caution should be taken regarding the overuse of postprocessing functions?
It enhances image resolution
It may improve diagnostic quality
It reduces the need for radiologist interpretation
It could drastically alter the original data set
21
Level 3 QC is labeled as:
Complex and noninvasive
Invasive and simple
Invasive and complex
Simple and invasive
22
First person to systematically collect and analyze patient outcomes and recommended changes based on those observations
Florence Nightingale
Ernest Codman
W. Edward Deming
Joseph Juran
23
The xray tube must be mounted so that the central ray exits the housing to within ____ degree of perpendicular.
2
4
1
3
24
Which term describes the following
Pixel averaging
Interpolation
Automatic Rescaling
Ghosting
25
__ is the Ability of imaging system to display events occurring close together in time as separate events
Visual resolution
Contrast Resolution
Temporal Resolution
Spatial Resolution
26
What two methods are used for measure image brightness?
Spinning top test/wire mesh tool
Pixel value readout/Photometer
Photometer/Oscilloscope
Ion chamber/Density monitor
27
Brightness and image contrast in digital imaging can be influenced by,
1 only
2 and 3 only
1 and 2 only
1, 2, and 3
28
This is an advantage of digital image storage:
reduced need for physical storage space
Reduced speed of image retrieval
increased image retrieval time
increased need for physical storage space
29
The minimum filtration required for general-purpose radiographic systems is _____ mm Al.
3.0
2.1
2.0
2.5
30
In mobile fluoroscopy (C-Arm), it is best practice to orient the tube in this fashion: (select 2)
1 and 5
2 and 6
3 and 4
2 and 4
31
Components of PACS include: Display workstations, Storage archive, ________, __________
LAN ; VPN
Network ; Image acquisition device
DICOM ; HL7
Router ; Network interface card
32
Which devices can be used as a sensor for AEC units
1
2
1, 2, and 3
1 and 2 only
33
The window width setting in a digital imaging system is used to manipulate which of the following values
Temporal Resolution
Spatial Resolution
Pixel brightness
Image Contrast
34
Which level of QC is performed by the technologist and requires no special training?
Level 2
Level 1
Level 3
35
PACS stores what type of image data
IHE images
Visible light images
Endoscopy images
DICOM images
36
For a xray unit operating at 80 kVp, _____ mm Al is the accepted HVL
2.5
2.3
2.1
2.7
37
In a DR flat panel, the DEL (detector element) size determines which of the following:
Contrast Resolution
Spatial Resolution
Detective Quantum Efficiency (DQE)
Dynamic Range
38
Collimation systems equipped with positive beam limitation must be accurate to within ____ of the SID
2%
3%
5%
1%
39
A pictorial representation of a frequency distribution
Trend chart
Fish bone chart
Scatter plot
Histogram
40
What type of artifact is this?
Dropped/dead pixels
Dirt/Scratch on plate
Moire Effect
Electronic interference
41
A computer ___ connects 3 or more computers that share resources and information
Node
Group
Hive
Network
42
Which is the proper acronym for: Digital Imaging and Communications in Medicine
DGCM
DIGCOM
DICM
DICOM
43
The system's ability to produce a quality radiograph despite
minor
errors in technical factors
Histogram analysis
Automatic processing
Histogram Repurposing
Automatic Rescaling
44
Patient motion during exposure results in ___ artifact
exposure
handling
no
processing
45
Most performance testing performed with gas-filled chamber detectors. The MOST often type of detector used is the ____
Ion Chamber
Proportional Counter
Geiger-Muller Counter
Gas Particle Detector Chamber
46
Which of the following are ADVANTAGES of Repeat Analysis:
2, 3, 4, and 5
1, 2, 4, and 5
1, 3, 4, and 5
1, 2, 3, 4, and 5
47
Image file size is the product of the ________ and the grayscale bit depth of the image
Image brightness
Spatial ratio
Temporal resolution
Matrix size
48
ensures the x-ray generator produces the same intensity of radiation each time the same set of technical factors is used
Linearity
Reproducibility
Beam quantity
Reciprocity
49
Detective quantum efficiency (DQE) is a measure of which of the following?
X-ray absorption
Contrast resolution
Image noise
Spatial Resolution
50
The DICOM image header does not specify an image’s
matrix size
modality
image acquisition parameters
interpretation time
51
What does MIMPS stand for?
Mammographic Image Management and Picture System
Medical Immediate Management and Pixel System
Medical Image Management and Processing System
Medical Information Messaging and Picture Statistics
52
Which type of testing requires fluoro-ing different thicknesses of material and seeing if brightness level is constant
kVp accuracy
AERC
Voltage waveform
Output Phosphor accuracy
53
What type of artifact is this?
Double Exposure
Grid cut off
Moire effect
Dropped/Dead pixels
54
In fluoroscopic units, the heat sensor alarm should activated when _________% of maximum heat is achieved.
25
10
80
75
55
What is the purpose of the DICOM standard?
to ensure interoperability of different imaging devices
manages patient procedure scheduling
manages financial transactions in healthcare
standardizes electronic health records
56
an Xray unit's timer must be within _____% accuracy for exposures less then 10 ms
20%
15%
5%
10%
57
Which level of QC is performed by the technologist and requires special training?
Level 3
Level 2
Level 1
58
When performing quality control tests on grids, the common tests are:
ratio/alignment
ratio/frequency
uniformity/ratio
alignment/uniformity
59
What is the term for: the average set of observations
mode
mean
variance
median
60
The backup system for an AEC should terminate the exposure at ___ second(s) or ___ mAs, whichever comes firs
1; 100
3; 300
5; 500
6; 600
61
Continuing Quality Improvement involves all employees
True
False
62
When testing the kVp, the actual kVp produced must be within ____ of the indicator on the control panel
10%
15%
2%
5%
63
This is a test tool for focal spot size
Spin top test
Star or slit pattern
Ion chamber
wire mesh test
64
What type of data is described as being measurable and focused on numbers and frequencies?
Quantitative
Perceived
Expected
Qualitative
65
If the tube type intensifier is in the presence of a strong magnetic field this type of distortion will appear
S distortion
Vignetting
Veiling glare
Pin cushion
66
Which type of test tool may be used to demonstrate pincushion distortion?
Collimator test tool
Oscilloscope
Wire mesh tool
Ion Chamber
67
In statistics, a sample is ___
frequency of occurrences
a pattern or trend of collected data
number of items actually measured from a population
the average measurements of encounters
68
Distribution of tissue densities and/or physiologic changes that are present in the anatomic part during the study
Dynamic contrast
Subject contrast
Radiant Contrast
Luminous Contrast
69
If a CR receptor hasn't been used for 48hrs, it should be inserted into the processor for erasure
False
True
70
Minimum source to skin distance for mobile fluoroscopic units is:
30 inches
8 inches
15 inches
12 inches
71
This is the type of quality is based the customer’s perception of the product or service.
Expected quality
Actual quality
Perceived quality
72
In digital imaging, as the size of the image matrix increases,
2 and 3 only
1 and 2 only
1 only
1, 2, and 3
73
What artifact will not interfere with an accurate diagnosis/interpretation of an image
Extreme quantum mottle
Moire Effect
A single dead row/column of pixels
Grid cut-off
74
______ represent a more comprehensive digital version of a patient's health information compared to _______
IHE ; HL7
EHR ; EMR
EMR ; EHR
RIS ; HIS
75
What do you do when there is quantum mottle on the fluoroscopic image?
increase kVp
increase mA
Instruct the patient to remain still for exam
Increase pulse rate
76
Which type of fluoroscopic QC test involves visualizing a MINIMUM mesh size on an image?
Distortion
High contrast resolution
Low contrast resolution
Focal spot accuracy
77
What is the reason for evaluating focal spot size & detecting any blooming during QC checks of an X-ray tube?
To ensure that the X-ray generator is functioning properly
To reduce the exposure time required for each imaging procedure
To ensure the X-ray tube is producing enough radiation for imaging
To maintain the spatial resolution and image quality
78
What artifact is this?
Moire effect
Grid cut-off
Dropped/dead pixels
Electronic Interference
79
_ compression is not reversible & can introduce some degree of data loss but can compress data size from 5 to 50 times
Lossless
Lossy
80
Which of the following terms is the unit most commonly used to measure luminance?
Lux
Foot-candle
Nit
Lumens
81
______ covers a large geographical area, functions at a slower rate, and support teleradiology.
NAS
LAN
DAS
WAN
82
____ is a device that emits electrons
Input phosphor
Photocathode
Output phosphor
Anode
83
Digital images are made of ____________, arranged in a ___________.
Pixels; matrix
DELs; matrix
Matrix; pixels
DELS; pixels
84
The maximum variation coefficient for reproducibility of x-ray generators is:
0.20
0.10
0.05
0.07
85
This is the quality level expected by the customer; influenced by word of mouth from others
Actual quality
Expected quality
Perceived quality
86
__ is categorized by use. Uses can be primary reading stations, QC stations, review stations for referring physicians
Acquisition modality
Archives
Mainframes
Display Workstations
87
The process of remote transmission and viewing of medical images is known as __________.
RIS
Teleradiology
PACS
DICOM
88
Fewer than _____distinct collimated margins may result in an image artifact (histogram error).
4
3
1
2
89
Which of the following can occur as a result of a double set of grid lines?
Grid misalignment
Static artifacts
Film fog
Moire effect
90
Proper collimation and ______________ prevent histogram errors that can lead to artifacts.
Algorithm
Technical factors
IR size
Centering
91
The range between the largest signal intensity, divided by the smallest signal value that a system can process/display
Automatic Range
Window Display
Dynamic Range
Histogram Range
92
What type of artifact is this?
Electronic Interference
Moire Effect
Dropped pixels/Dead pixels
Grid cut-off
93
This technique allows the transmission of large amounts of image data to remote locations (telerad) for interpretation
Flatfielding
Compression
Minification
Reduction
94
______________ contains patient imaging history, procedure scheduling information and the radiologists report.
HL7
HIS
EMR
RIS
95
The central part of PACS; houses all historic and current medical image data being generated in a healthcare enterprise
Archive
Display workstation
Acquisition modality
Clients
96
X-ray intensity of a beam is decreased by 50% when 4 mm aluminum equivelant filtration is added. This describes:
ABC
EHR
HVL
AERC
97
What is a Vendor Neutral Archive, in the context of medical imaging?
Archive that only supports Dental images
Archive that only support DICOM images
Archive that support only Pathology images
Archive that can store digital images like DICOM, Dental, Pathology images
98
Level 1 QC is Noninvasive and simple
True
False
99
___ customers come from outside the healthcare organization
internal