The magnification factor is 1.5 on an x-ray, & a mass on the x-ray measures 12cm, what is the actual mass size?
8cm
6cm
4cm
2
In order to maintain 45” SID, when the tube is angled 35°, the new vertical distance should be 36 inches.
False
True
3
Using the air‑gap technique, if the anatomy is 10 cm from the image receptor, the SID should be
approximately
:
40"
20"
30"
50"
4
Patient exposure is reduced with the use of speed numbers above 400 as compared to those speed numbers below 400.
True
False
5
With digital imaging,
quantum
mottle
will be demonstrated on underexposed CR images
True
False
6
With digital imaging, as the number of pixels increase, the image resolution decreases.
False
True
7
As the kVp level is decreased, does the amount of photoelectric interaction
increase OR decrease?
Decrease
Increase
8
What would happen to the anode if the temperature was
rapidly
increased
without the use of the warm-up technique?
Pitting of the anode
Bullet like melt
Cracking the anode
Filament Break
9
On the Radiographic Rating chart, if the intersection of kVp & exp. time is
below the mA curve
, the exposure is
unsafe
.
False
True
10
The total brightness gain (approx. 8-25k) refers to increased light emission from:
output phosphor
electrostatic lens
photocathode
input phosphor
11
The percentage of x‑ray photons scattered is greater at 50 kVp than at 100 kVp.
True
False
12
In digital radiography, the bit system is used to describe the:
dynamic range
overall image density
number of pixels in the image matrix
CR imaging technical factors
13
What is the purpose of the ADC in digital radiography?
Convert the electronic signal to a digital value
Adjust the brightness of the fluoro image
Convert the digital signal to an electronic signal
14
Of the following sets of exposure factors, which one will produce the
highest image
contrast?
80kVp 100mAs 40"SID 8:1 grid
50kVp 100mAs 40"SID 12:1 grid
90kVp 100mAs 40"SID 5:1 grid
70kVp 100mAs 40"SID 8:1 grid
15
If the kVp is decreased, does the contrast on the image: i
ncrease OR decrease?
Decrease
Increase
16
Off-focus (stem) radiation originates at the:
collimator
patient
cathode
anode
17
Which type of phosphor is used in CR photostimulable plates?
amorphous selenium
barium fluorohalide
calcium tungstate
18
When selecting the small filament, the effective focal spot size would be:
0.3 - 1.0 mm
0.1 - 0.3 mm
1.0 - 1.3 mm
1.0 - 3.0 mm
19
The size of the
image matrix
controls the:
image noise
size of radiographic image
patient dose
image resolution
20
A 14” x 17” of the abdomen requires 20 mAs @ 80 kVp. If the field is coned to an 8” x 10” what mAs should be used?
50 mAs
40 mas
20 mAs
30 mAs
21
During tomography, as the exposure arc
increases
,
does the detail on the image:
increase OR decrease?
Decrease
Increase
22
Size distortion on an x‑ray image results from:
elongation and foreshortening
a large focal spot
a large space between the object and the image receptor
unequal exposure to the object
23
The “photostimulated luminescent” process that occurs in the digital reader/processor results from:
collection of electrons on the positive electrode
conversion of the electronic signal to a numerical value
stimulation of electrons to release light & drop back into orbit
conversion of light to an electronic signal
24
According to Federal Regulations, the length of the exposure cord on a portable x-ray unit must be 5 feet.
False
True
25
The “anatomical image pattern” emitted by the photocathode is formed by:
x-ray photons
light photons
electrons
laser scanner
26
With tomographic imaging, an arc of greater than 10 degrees is referred to as zonography.
True
False
27
Disproportional magnification will occur when doing a PA axial projection of the skull.
False
True
28
In digital imaging, if the pixel has a
high numerical
value, that would be consistent with a thinner anatomical part.
True
False
29
A fiberglass cast requires what technical adjustment as compared to a non-casted extremity?
no change
increase the kvp by 10
decrease the kVp by 5
increase the kVp by 15
30
Frequency describes: The distance from a point on one wave to the same point on the next wave.
False
True
31
With digital imaging, underexposed CR images will demonstrate
low contrast.
True
False
32
The fuzzy, out-of-focus border of the anatomical part is referred to as umbra.
False
True
33
Tomography equipment is used for producing in‑focus plane images
False
True
34
When using the Agfa CR System, properly exposed imaging plates should produce an overall “Lgm” value of:
1.85 - 2.5
1.85 - 4.0
1.0 - 2.0
1.0 - 3.0
35
The half value layer is used to measure x-ray beam:
quality
quantity
36
Exposure latitude is controlled by the kilovoltage selection.
True
False
37
Geometric unsharpness can be controlled by increasing the OID.
True
False
38
Phantom images seen on tomographic images are: false images, or other blurred structures
False images
Blurred structures
39
The percentage of x‑ray photons scattered is greater at 50 kVp than at 100 kVp.
False
True
40
An adult KUB technique is 40 mAs, 80 kVp. For a 4‑year old child, you could therefore use 30 mAs, 75 kVp.
True
False
41
The image intensifier's
input
phosphor
is generally composed of:
zinc cadmium sulfide
barium lead sulfate
cesium iodide
calcium tungstate
42
Sharpness in detail on an image is dependent on mAs.
False
True
43
Inherent filtration consists of:
The glass tube
B & C
The housing oil
Sheets of aluminum
44
Acceptable techniques for 3‑phase equip. must be ____ if doing that same procedure using single‑phase equip.
decreased
increased
no change
45
To maintain the image density, if the kVp is decreased by 15%, how much should the mAs be increased?
8%
100%
15%
50%
46
Starting & stopping the rotor switch
unnecessarily
will eventually result in what type of tube failure?
Bullet like melt
Pitting of the anode
Cracking the anode
Filament Break
47
The intensity of the x-ray beam describes the x-ray beam’s penetration ability.
False
True
48
A fluoro exam is done using 120 kVp, 3 mA, and takes 3 minutes. The amount of anode heat units accumulated would be:
56,000 HU
92,000 HU
78,000 HU
36,000 HU
49
What is the formula used to determine the total heat units for high frequency x-ray equipment?
kVp x mA x time x # of exposures x 1.4
kVp x mA
kVp x mA x time
kVp x mA x time x # of exposures
50
To prevent excessive exposure to the patient, the fluoro lead shutters should never exceed the size of the:
output phosphor
input phosphor
patient’s anatomy of interest