Clinical Chemistry section 2 Flashcards | StudBud
Explore
/
Clinical Chemistry section 2
Clinical Chemistry section 2
83 terms
·
StudBud Administration
·
22 plays
Imported from external platform
Flashcards
83 cards
F
Practice quiz
P
Host live
L
Questions
83
Answers
1
_____ should be tested within 4 hours or stored refrigerated/frozen
Serum or Plasma
Urine
Saliva
Tissue
2
reject when 1 control measurement in a group exceeds the mean +2sd, and another exceeds the mean –2sd.
R4s
12s
22s
13s
3
This rule identifies systematic error only and can be applied within a run or across runs.
22s
12s
R4s
13s
4
These types of errors are unpredictable and widen the distribution of values.
Sytematic
Random
5
controls may be purchased with value ranges for tests of interest
unassayed
assayed
6
Speed of approximately 3000-3500 rpms (centrifugal force) for _______ is necessary to remove cells
3-5 minutes
24 hours
10-15 minutes
1-5 minutes
7
allowable error is
best if greater than the total error
published by the CDC
used only for research purposes
a combination of random and systemic error
8
The likelihood of a negative test result in patients without disease; it measures how well the test excludes disease
Diagnostic Sensitivity
Diagnostic specificity
Medical Decision Limits
analytical sensitivity
9
Testing that is used to diagnose disease should detect all positives without giving any false positives.
Waived Test
Screening Test
Confirmatory Test
Diagnostic Test
10
Measurement of unknown samples that are sent from an outside agency at set intervals, as a test of competency.
reference range
reliability
external quality control/proficiency testing
standard deviation
11
A Gaussian distribution is usually
bell Shaped
uniform
skewed
scattered
12
For a data set that is Gaussian distributed, what's the likelyhood that a data point will be within ±1 SD from the mean?
68%
57%
99%
95%
13
An initial test that should detect everyone with the disease before they present with symptoms
Confirmatory Test
Diagnostic Test
Waived Test
Screening Test
14
is the ability to detect small concentrations of the measured analyte
reliability
analytical sensitivity
analytical specificity
relevancy
15
is freedom from interference and cross reactivity, to determine solely the analyte that is being measured
analytical sensitivity
relevancy
analytical specificity
reliability
16
SOP stands for Standard Operating Procedure
True
False
17
refers to everything that affects a patient specimen before it is tested for the analyte.
variance
analytical
post-analytical
pre-analytical
18
The _____ is the variable that falls halfway between the highest and the lowest in a group of numbers.
Mode
Median
Bias
Mean
19
Advantages of gel tubes
do not
include:
Lower frequency of error with testing
Less manipulation of the specimen
preserving glucose for up to 5 days
Less chemical interaction of blood to serum
20
The test method that provides the best available approximation of a true value.
The Gold Standard Method
confirmatory screening
recovery studies
interference studies
21
Due to concentration, the first urine of the morning is optimal for testing.
True
False
22
Computed values that compare your lab with other labs that are using the same instrument with the same methodology.
Standard Deviation Index
Westgard Rules
Standard Operating Procedure
Levey Jennings Chart
23
This rule identifies unacceptable random error or possibly the beginning of a large systematic error
13s
R4s
12s
22s
24
The range of values for a given analyte in a defined population
Gaussian Curve (Bell-shaped Curve)
Reference Range
Standard Deviation Index (SDI)
Bias
25
____ is concerned with sensitivity, specificity, and closest to the true value
precision
trend
accuracy
shift
26
At the manufacturer samples are checked against analytical methods known to be free of interference.
confirmatory screening
interference studies
recovery studies
comparison testing
27
Green Top/PST are used for chemistries and contain:
Clot activator, no anticoagulant
Sodium fluoride and potassium oxalate
Sodium heparin, Lithium heparin, Ammonia heparin
Sodium EDTA
28
Serum or plasma should be separated within
2 days
1 hour
24 hours
2 hours
29
Delta checks are particularly helpful in detecting misidentified specimens.
False
True
30
A test used to verify the results of an initial test.
Screening Test
Waived Test
Confirmatory Test
Diagnostic Test
31
A number that describes the number of standard deviations away from the mean that a result is.
Bias
Standard Deviation
Coefficient of Variation (CV)
Standard Deviation Index (SDI)
32
Dark blue tops are used for testing trace elements and this additive within them forms calcium salts:
Clot activator, no anticoagulant
Sodium EDTA
Sodium fluoride and potassium oxalate
Sodium heparin, Lithium heparin, Ammonia heparin
33
In general, a CV of <10% is acceptable, and for low level assays, a CV of <15% is acceptable.
False
True
34
_____ is reproducibility of results.
trend
accuracy
precision
shift
35
The greatest percentage of laboratory errors occurs during this phase:
proficiency testing
post-analytical
analytical
pre-analytical
36
a gradual but steady change in quality control results
relevancy
trend
shift
precision
37
A statistic that describes the distribution of all values around the mean for a particular group.
Standard Deviation
Coefficient of Variation (CV)
Bias
Standard Deviation Index (SDI)
38
Analytical activities include: specimen & reagent preparation, instrument calibrations, and providing the lab report.
False
True
39
QC samples are produced from pooled serum samples. Purchased controls may be assayed or unassayed.
False
True
40
The clinical chemist is concerned with two aspects of information: Accuracy and Relevance
True
False
41
A reference range can be verified by:
testing as few as 20 normal donor specimens
samples from the pharmacy
CLIA
literature and vender review material
42
Accurate time and patient compliance
are not
important factors in collecting urine specimens.
True
False
43
A sudden and sustained change.
shift
precision
relevancy
trend
44
The range of specimen concentration that can be accurately measured by a test method
reportable range/analytical measurement range/linear range
total error
patient comparison
allowable error
45
Patient position, diet, pregnancy, and stress levels have no affect on testing results.
False
True
46
Systematic error can be best described as consisting of:
total errors
syntax errors
random errors
constant and proportional error
47
______ evaluate individual test results on the basis of normal variation.
Proficiency Testing
Standard Operating Procedures
Standard Deviation Index
Delta Checks
48
There are 4 major legitimate reasons for ordering a laboratory test: diagnosis, monitoring, research and screening
True
False
49
Proficiency testing consists of _____ challenges _____ times a year for each test that the laboratory offers.
5, 3
3,4
3,5
5, 4
50
are cutoff values that help differentiate the presence of a particular disease from the absence of that disease.
reference range
analytical sensitivity
Medical Decision Limits
analytical specificity
51
24 hour urine collection reduces diurnal variation
True
False
52
reject when 2 consecutive control observations on the same side of the mean exceed the mean ±2sd limit
22s
R4s
12s
13s
53
The achievement of both precision and accuracy.
trend
reliability
relevancy
precision
54
refers to those variables that affect patient results in the reporting stage
variance
pre-analytical
analytical
post-analytical
55
refers to every step involved in the actual analysis of a laboratory specimen.
post-analytical
analytical
variance
pre-analytical
56
Post-analytical activities do not include:
Reporting patient results in a timely manner
Notifying health care providers when results arrive from outside reference
Positively identifying the test report with the patient sample.
Collection of specimen
57
QC monitors the following aspects of testing:
Technical error by personnel
All of these
Reagents and other testing products and equipment
Performance of laboratory instruments
58
The Westgard rules use a combination of decision criteria, to decide whether an analytical run is in or out of control.
True
False
59
This rule warns that random error or systematic error may be present in the test system
12s
22s
R4s
13s
60
Ideally, the amount of the standard added should be recovered at the end of the measurement if the method is accurate.
confirmatory screening
comparison testing
interference studies
recovery studies
61
Pre-analytical activities include: collection, labeling, storage, processing, and transportation of specimen.
False
True
62
when 1 control observation exceeds mean +/- 2sd
12s
R4s
22s
13s
63
Red Top/ Tiger Top SST/Gold SST are used for chemistry testing and contain:
Sodium heparin, Lithium heparin, Ammonia heparin
Sodium EDTA
Sodium fluoride and potassium oxalate
Clot activator, no anticoagulant
64
_______ is/are the most common method used in a clinical laboratory to plot quality control results.
A Levey-Jennings chart
Standard Deviations
The Westgard Rules
Reference Ranges
65
is a reference point to determine whether a disease is present or absent, or if a patient is at risk for future disease
reference range
analytical specificity
analytical sensitivity
relevancy
66
Control ranges for _______ controls must be calculated
unassayed
assayed
67
Gold standard methods have been established and reproduced across different measurement systems and laboratories.
False
True
68
Used to compare the standard deviations of two samples. Can be used to compare test results within the same lab.
Standard Deviation Index (SDI)
Coefficient of Variation (CV)
Standard Deviation
Bias
69
Balancing the centrifuge isn't important at all.
True
False
70
______is the amount of error that can be tolerated without invalidating the medical usefulness of the analytical result
allowable error
sensitivity
total error
standard deviation
71
is the
average
of a set of numbers
Mode
Bias
Mean
Median
72
Light gray tops are used for glucose testing and this additive helps preserve glucose for up to 5 days:
Clot activator, no anticoagulant
Sodium heparin, Lithium heparin, Ammonia heparin
Sodium EDTA
Sodium fluoride and potassium oxalate
73
Interference studies typically use _____ as an interference
All of these
hemolyzed red cells
highly lipemia samples
highly icteric samples
74
This type of error causes a shift in the mean of the distribution
Sytematic
Random
75
The likelihood of a positive test result in patients with disease; it measures how well the test detects the disease.
Diagnostic Sensitivity
Medical Decision Limits
Diagnostic specificity
analytical sensitivity
76
Which color tube isn't a common tube for blood chemistry testing?
Lavender -contains EDTA
Red top –no additives (glass – no clot activator, plastic – clot activator
Red/Gray SST – clot activator and gel serum separator
Gold tube - may or may not have clot activator, have gel serum separator
77
QC material is a solution or chemical substance of known concentration that mimics a patient’s specimen.
True
False
78
reject when 1 control observation exceeds mean +/- 3sd
13s
R4s
12s
22s
79
This rule identifies random error only and can be applied to one control level on consecutive days
12s
13s
22s
R4s
80
The ______ is the value that occurs most commonly in a mass of data
Bias
Mode
Mean
Median
81
is the ability to detect small concentrations of the measured analyte
reliability
analytical specificity
analytical sensitivity
relevancy
82
_______ is the sum of random and systematic error
standard deviation
allowable error
sensitivity
total error
83
is a process that includes a set of activities that aids in improving the reliability of the laboratory as a whole.
Quality Assessment/Assurance/Management
Accuracy
Quality Control
Precision